Intsholongwane yeHepatitis B

Inkcazo emfutshane:

Le khithi isetyenziselwa ukufumanisa ubungakanani be-hepatitis B virus nucleic acid kwiisampuli ze-serum yabantu kwi-in vitro.


Iinkcukacha zeMveliso

Iithegi zeMveliso

Igama lemveliso

Ikhithi yokuFumanisa iNucleic Acid yeHWTS-HP001-Hepatitis B Virus (iFluorescence PCR)

I-Epidemiology

I-Hepatitis B sisifo esosulelayo esinezilonda zesibindi kunye namalungu amaninzi abangelwa yintsholongwane ye-hepatitis B (HBV). Uninzi lwabantu lufumana iimpawu ezifana nokudinwa kakhulu, ukuphelelwa ngumdla wokutya, amalungu asezantsi okanye ukudumba komzimba wonke, i-hepatomegaly, njl. I-5% yabantu abadala kunye ne-95% yabantwana abagulayo abasuleleke ngoomama babo abakwazi ukuyicoca intsholongwane ye-HBV ngokufanelekileyo kwisifo esingapheliyo kwaye baqhubekele kwi-cirrhosis yesibindi okanye i-carcinoma yeseli yesibindi ephambili..

Itshaneli

I-FAM I-HBV-DNA
I-VIC (HEX) Isalathiso sangaphakathi

Iiparameters zobugcisa

Indawo yokugcina izinto ≤-18℃ Ebumnyameni
Beka ubomi kwishelufa Iinyanga ezili-12
Uhlobo lweSibonelelo Igazi lemithambo yegazi
Ct ≤33
CV ≤5.0%
I-LoD 25IU/mL

Ukuchaneka

Akukho kudibana kwe-Cytomegalovirus, i-EB virus, i-HIV, i-HAV, i-Syphilis, i-Human Herpesvirus-6, i-HSV-1/2, i-Influenza A, i-Propionibacterium Acnes, i-Staphylococcus Aureus kunye ne-Candida albican.
Izixhobo Ezisebenzayo Ingalingana nezixhobo zePCR ezikhanyayo ezithengiswayo.

Iinkqubo ze-ABI 7500 ze-Real-Time PCR

Iinkqubo ze-PCR ze-ABI 7500 ezikhawulezayo zeXesha langempela

Iinkqubo ze-PCR ze-SLAN-96P zeXesha langempela

Iinkqubo ze-PCR zeQuantStudio®5 zeXesha langempela

Iinkqubo ze-PCR zeLightCycler®480 zeXesha langempela

Iinkqubo zokuFumanisa i-LineGene 9600 Plus zeXesha langempela ze-PCR

I-MA-6000 Real-Time Quantitative Thermal Cycler

Inkqubo ye-PCR ye-BioRad CFX96 yeXesha langempela

Inkqubo ye-PCR ye-BioRad CFX Opus 96 yeXesha langempela

Ii-reagents zokukhupha ezicetyiswayo: Macro kunye neMicro-TestIntsholongwaneI-DNA/RNA Kit (HWTS-3017) (engasetyenziswa ne-Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-EQ011)) yiJiangsu Macro & Micro-Test Med-Tech Co., Ltd.. Ukukhupha kufuneka kuqalwe ngokwe-IFU ye-reagent yokukhupha. Umthamo wesampulu okhutshiweyo yi-200µL kwaye umthamo wokucoca ocetyiswayo yi-80 μL.

Ii-reagents zokukhupha ezicetyiswayo: Ii-Nucleic Acid Extraction okanye ii-Purification Reagents (YDP315). Ukukhupha kufuneka kuqalwe ngokuhambelana ne-IFU. Umthamo wesampulu okhutshiweyo yi-200µL kwaye umthamo wokucoca ocetyiswayo yi-100 μL.